SYRINGE : 10, 20, 30, 50/60ML standard with automatic size detection, 5ML*, 100ML *, Front loading of syringe, All available standard brands of syringes. Custom Programmable to any brand syringe.
FLOW RATE RANGE : 0.1-999.9ml /Hr (Depending on syringe size) with increment of 0.1mL/Hr. Higher rate available on request
: Increment of 0.01mL/Hr for range up to 99.99mL/Hr *
INFUSED VOLUME RANGE : 0.1 -999.9ml
BOLUS RATE & VOLUME : 300-999.9ml/Hr (Depending on syringe size), Volume is programmable by user
: Higher rate (>1200mL/Hr) & programmable Bolus Rate available on request.
BOLUS INTERVAL : A minimum interval of time can be preset between two Bolus dose to avoid accidental over infusion, 1min to 24 Hrs.
PURGE FACILITY KVO : To Prevent Needle Clogging with Programmable Kvo Rate
OCCLUSION : 3-4 Levels. Anti bolus mechanism in case of occlusion
: 9 Levels or more *
: Monitoring of Actual In Line Pressure & Set Occlusion Level *
PUMP STACKING : Multiple Pump Stacking with Normal or Optional Single Power Cord.
: Facility of Stacking of 2-3 Pumps with built in Stacking Clips, ensures more useful group application & portability in & around the
hospital premises.
: Integrated pole clamp
BODY WEIGHT MODE * : Delivery rate is calculated automatically based on Volume & time pre selection or on values entered for patient body weight with drug
parameters.
DRUG LIBRARY : 48 Standard Drugs
: More No. of Drug names available on request.
MULTI STAGE PROGRAMMING * : Ramp up & Ramp down facility
ALARMS : Occlusion, End of syringe, Low Battery, Syringe displaced, No Syringe, Wrong Syringe, Pre Alarm for End, Invalid rate, Target
Volume, Target Time, Target volume Pre-alarm, Mains Failure Alarm, Preventive Maintenance Alarm.
: Line Disconnection Alarm *
ACCURACY : Volume +/- 1%, Time +/- 1%
TIME LIMIT : Time from 00:01 to 23:59 (hrs:min) or more if specially demanded
VOLUME LIMITS : From 0.1ml to capacity of the syringe
: Higher volume limits on request.
POWER : 220 VAC, +/- 10%, 50/60 HZ 12VA
: External DC Source *
BATTERY TYPE : Ni Mh, 12V/9.6V, 1300 mAHr
BATTERY BACK UP : 5-6 hrs (Typ.) @ 5ml / Hr. Or more if specially demanded
BATTERY RECHARGING : 20 Hrs. For full charging
STANDARDS : Tested as per IEC 60601.1, 60601-1-2, 60601-2-24
DISPLAY : Premium look with Green Color Display
: Exclusive Blue Color Display (Only in Infutek 405Ex)
RS485/RS232 PORT : For Computer Interface Central Monitoring Station (Only in Infutek 405Ex)